Fluid Quantities
Burettes
| Parameter / Equipment |
| Burette (Buret) Calibration |
Refractometers
| Parameter / Equipment | Range | CMC Uncertainty (±) |
| Refractometer Calibration | 1.33659 nD 1.34401 nD 1.35568 nD 1.46990 nD 1.52256 nD |
0.00042 nD 0.00042 nD 0.00042 nD 0.00048 nD 0.00049 nD |
Viscosity Meters
| Parameter / Equipment | Range | CMC Uncertainty (±) |
| Viscosity Meter (Viscometer) Calibration | Up to 33 % of Scale (33 to 66) % of Scale (66 to 100) % of Scale |
2.7 % of rdg 1.8 % of rdg 1.7 % of rdg |
Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.
Calibration
and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 %
level of confidence, usually using a coverage factor of k = 2. The actual measurement uncertainty of a
specific calibration performed by the laboratory may be greater than the CMC due to the behavior of
the customer’s device and to influences from the circumstances of the specific calibration.
Please note the actual measurement uncertainties achievable on a customer's site can normally be expected to be larger than the CMC found on the A2LA Scope. Allowance must be made for aspects such as the environment at the place of calibration and for other possible adverse effects such as those caused by transportation of the calibration equipment. The usual allowance for the actual uncertainty introduced by the item being calibrated, (e.g. resolution) must also be considered and this, on its own, could result in the actual measurement uncertainty achievable on a customer’s site being larger than the CMC.